肝内の周波数信号を利用した体表組織の減衰値の推定法-脂肪肝の肝エコーレベル測定への応用-
Measurement of Ultrasonic Attenuation Loss in Surface Wall Using Reflecting Signal in the Liver and Estimation of Echo Level of the Fatty Liver
谷口 信行, 伊東 紘一, 中澤 成公, 中村 みちる, 森 博昭
Nobuyuki TANIGUCHI, Kouichi ITOH, Yoshitaka NAKAZAWA, Michiru NAKAMURA, Hiroaki MORI
自治医科大学臨床病理学教室
Department of Clinical Pathology Jichi Medical School
キーワード : Echo level, Fatty liver, Attenuation, Thorax, Ultrasound
Because of the difficulty of measuring attenuation in skin and subcutaneous tissue in vivo, we estimated attenuation indirectly by using mean frequency shifts in the liver. This method can readily measure attenuation by correlating depth and mean frequency shift in the liver. The attenuation coefficient (mean±standard deviation) in the right intercostal wall was 1.69±0.61 dB/cm/MHz (n=23); and the wall thickness was 1.9±0.1 cm. Attenuation was inversely correlates with wall thickness. The strength of the liver echo is estimated by using an echo-intensity histogram. The echo level of the liver is then corrected by compensating for thoracic wall attenuation. This method was used on the control group and fatty-liver group. The echo level in the fatty-liver group was high only after it was corrected for wall attenuation, compared with the control group.